EFFICIENT USE OF VALUABLE SPACE
The most popular model in the HettCube incubator range.
Building on the base of the 400 model, the HettCube 600 has a maximum validated usuable space of 351 l.
This model is capable of holding up to 1,800 petri dishes.
This unit is available with refrigeration and a temperature range from 0 °C to +65 °C (HettCube 600 R).
Highlights
Features
Technical Data
- maximal validated usable space on a small footprint
- One-hand operation without additonal inner door
- 4.3 inch touch displayfor intuitve operation
- automatic door closure with magnetic seals, door orientation can be easily changed
- low noise level of ≤ 44 dB(A)
- TÜV certified
- Factory certificate HettCert with 9 measuring points according to DIN 12880:2007-5 standard
- HettCube without cooling requires only 1°K (1 °C) above stable ambient temperature
- No interior fan
- Model variants with IVD approval incluiding DIN 62304 standard
- Inclusive: 3 standard shelves and 1 HTS shelf with telescopic rail Additionally for cooled models:
- Temperature drop with holiday function
- Temperature selection monitor class 3.2
- Easy integration of external devices via the switchboard (optional)
- At a glance display of: settings, future programs and target/actual values, temperaure drops, events
- real-time calender
- More individual settings: Flexible door alarm, asymmetric tolerance band
- Three different power failure scenarios
- Logbook: records of every door opening, events, alarms and operating hours
- Language options EN, DE, FR, ES
- Wide range of program functions: Start after time, start after temperature, standstill segment
- Programs and temperature settings transferable to other HettCubes without additional software
- Interior made of high-quality stainless steel Special accessories:
- Shelfs and racks for storage of samples
- shelf for Petri Dishes
- tightly welded HTS drawers (various heights)
HettCube 600 incubators | HettCube 600 R cooled incubators | |
Temperature range | 1 K above ambient temperature up to +65 °C | 0 °C up to +65 °C |
Exterior dimensions (without bushing and door handle) W x D x H in mm | 710 x 825 x 1990 | 710 x 825 x 1990 |
Interior dimensions W x D x H in mm | 535 x 690 x 1415 | 535 x 690 x 1415 |
Internal volume in liters | 520 | 520 |
Validated usable volume in liters | 351 | 351 |
Percentage validated usable volume / internal volume | 67 % | 67 % |
Footprint in m² | 67 % | 67 % |
Weight in kg | 164 | 175 |
Number of trays provided as standard | 4 (3 standard + 1 HTS) | 4 (3 standard + 1 HTS) |
Temperature fluctuation at +37 °C | ± 0.1 K | ± 0.1 K |
Temperature uniformity at +37 °C | ± 0.2 K | ± 0.2 K |
Temperature uniformity at +25 °C | ± 0.1 K | ± 0.1 K |
Recovery time after door has been opened for 30 s at +37 °C | ≤ 5,5 min | ≤ 5,5 min |
Energy consumption at +37 °C | 0.056 kWh | 0.056 kWh |
Noise level | ≤ 41 dB (A) | ≤ 44 dB (A) |
Power supply | 220 – 240 V 1 ~ / 50 – 60 Hz | 220 – 240 V 1 ~ / 50 – 60 Hz |
Cat. No. | 66000 | 66005 |
Non-IvD version | 66001 | 66006 |
120 V 1 ~ / 50 – 60 Hz | 66000-01 | 66000-01 |
Weight in kg* | 169 | 180 |
Energy consumption at +37 °C* | 0.06 kWh | 0.06 kWh |
Noise level* | ≤ 42 dB (A) | ≤ 42 dB (A) |