CytoFAST Elite Cytotoxic Safety Cabinets belong to the latest generation of laminar airflow systems manufactured by Faster, in which the choice of materials of construction of the highest quality guarantees conformity to the strictest safety standards.
CytoFAST Elite are Cytotoxic Drug and Microbiological Safety Cabinets designed and built to performance requirements of the EN 12469:2000 and DIN 12980:2005-06 European Standards. They are triple HEPA filter technology cabinet, with 100% of the air filtered via the main H14 HEPA/ULPA filter directly below the work surface and then 70% of the air re-circulated via the recirculating H14 HEPA/ULPA filter within the cabinet, whilst the remaining 30% is discharged through an exhaust H14 HEPA/ULPA filter.
CytoFAST Elite have been designed to guarantee the product, personnel and environmental protection and they are primary suitable for applications such as: preparation and handling of cytotoxic drugs, preparation and manipulation of antineoplastic chemotherapeutics, preparation and manipulation of CMR drugs and any other agents likely to cause genetic alterations or synergetic activities with other materials.
Since CytoFAST Elite meet also EN 12469 as for Class II Biohazard cabinets, they can be adopted also for a wide range of disciplines in applications such as: Microbiology, Virology, Haematology, Cell culture, Genetics, Handling of hazardous agents to human beings or animals.
The ambient air is drawn in from the slots at the stainless-steel base of the front opening and it then passes through the H14 HEPA/ULPA filter bank below work surface, from where it is drawn up and blown into the plenum of the re-circulating and exhaust fan.
The “bio-dynamic sealing system” of the negative pressure plenum ensures that all contaminated particles are kept inside the system and are automatically drawn to the plenum or pressure chamber to be captured by the main re-circulating and exhaust H 14 HEPA/ULPA filters.
The fan system assures that no part of the cabinet comes ever under positive contaminated pressure to the laboratory, thus protecting and preserving the environment and operating personnel from exposure to cytotoxic drug, CMR compounds and agents of bio-contamination.
70% of the filtered air is re-circulated (after passing through a H14 HEPA/ULPA filter) in an air cleanliness in Class ISO 3, according to ISO 14644-1 laminar flow pattern downwards into the work chamber and the remaining 30% is exhausted to atmosphere through another H14 HEPA/ULPA filter.
Ergonomic Design: The angled sloping (7°) front stratified safety-glass sash provides optimum visibility of all objects placed in the interior workspace. The sash is electrically operated. Pressing the appropriate touch-sensitive keys will completely open or completely close down the sash. The standard sash-height opening during work is set to 200 mm. Alternative sash-height settings (250-160mm) by the factory are possible upon request. The whole front of the safety cabinet – which includes the sash mechanism – can also be opened upwards as it is hinged on the top – in order to allow easy access for complete and effective cleaning and decontamination.
Real Laminar AirFlow: Frontal screen 7° sloped as well as back side wall to convey in unidirectional pattern the air flow. As a consequence, the front and back panels are parallel one with the other and there is the real presence of Laminar AirFlow in the whole working area.
Innovative system of third bank H 14 HEPA/ULPA filtration beneath the working surface with increased space for operator’s knees and comfort (>480mm): the filters are conceived with the suitable size to be disposed in a standard 60lt waste bin according to the latest version of DIN 12980:2005-06 Standard. Bag-Out System: Safe and easy H14 HEPA/ULPA filter replacement beneath the working surface for service technicians and environment protection according to DIN 12980:2005-06 Standard.
ECS® Eco Controling System: The new ECS® microprocessor employs the latest innovative methods of integrated management of all principal functions of ventilation and filtration – self-regulating all the main filtration and ventilation system components – compensating for declining pressure drops and restoring power balance. Combining the use of AC motor-blowers and certified low pressure-drop filters, the new ECS® controlling system optimize power consumption, reducing CO2 emissions into the environment.
Anti Bacterial Coating: Exclusive Dupont™ ALESTA® anti-bacterial “Ag+ cations-based solution”, capable to prevent microbial contamination of surfaces thereby inhibiting long term surface growth.
Inflatable gasketing is also available as an option to provide effective sealing of the internal workspace of the chamber for purposes of fumigation/sterilization and when gas-tight is needed.
Silent Operation: The TNT plenum, the structures of the electric motors of the fans fitted on their antivibration mounts and the software itself designed to provide optimum air handling characteristics guarantee quiet operation of this silent safety cabinet, with sound-pressure levels recorded way below the parameters specified in the current EN:12469 European Standard for Microbiological Safety Cabinets.
Protection grid: A protection grid made of AISI 304L stainless steel is provided inside the part behind the rear panel (where air flows), in order to avoid the tissue used during cleaning operations be sucked in the motor-blower causing the failure of the blower and other damages.
High Level Lighting: The safety glass side-windows with the ideal positioning and sizing of the light-system provide the highest level of luminosity to the work area.
Easy handling and maintenance: The cabinet can pass through standard 800 mm wide door openings. In fact, the overall depth of the cabinet can be reduced to approx. 777 mm by removal of the rear panel. All service operations are available from the front of the cabinet. Despite the overall height, CytoFAST Elite have a demountable base stand of 255 mm. Without this stand the cabinet height is 1995 mm and can easily pass through doors of 2 meters height.
Automatic safety service connections for gas and vacuum and one (for size 209 and 212) or two (for size 215 and 218) electrical socket(s) fitted as standard in each size model.
Code | F000217A0000 |
Width [cm] | 150 |
Pressure | Negative |
Motor-blowers | Double fan |
Useful dimension (mm):: | WxDxH: 1497x580x740 |
Overall dimension [mm]: | WxDxH: 1655x855x2345 |
Noise level (dbA): | <55 |
Working Aperture (mm): | 200 |
Exhaust flow rate (m3/h) | 485 |
Lighting level (lux): | >1200 |
Weight (kg): | 285 |
Electrical data [230V – 50H] (*) | 230 V – 50 Hz |
Current consumption (A) | 3,6 |
Heat emission (W) | 280 |