EFFICIENT USE OF VALUABLE SPACE
With its volume and wide range of accessories the HettCube 400 incubator is well suited for all pharmaceutical, research and industrial applications.
This unit is available with refrigeration and a temperature range from 0 °C to +65 °C (HettCube 400 R).
Please be aware that the HettCubes are currently available within the EU, Switzerland and Russia. Further countries will follow soon.
Highlights
Features
Technical Data
One-hand operation without additonal inner door
- 4.3 inch touch display for intuitve operation
- automatic door closure with magnetic seals, door orientation can be easily changed
- low noise level of ≤ 44 dB(A)
- TÜV certified
- Factory certificate HettCert with 9 measuring points according to DIN 12880:2007-5 standard
- HettCube without cooling requires only 1°K (1 °C) above stable ambient temperature
- No interior fan
- Model variants with IVD approval including DIN 62304 standard
- Inclusive: 2 standard shelves and 1 HTS shelf with telescopic rail Additionally for cooled models:
- Temperature drop with holiday function
- Temperature selection monitor class 3.2
- Easy integration of external devices via the switchboard (optional)
At a glance display of: settings, future programs and target/actual values, temperaure drops, events
- real-time calender
- More individual settings: Flexible door alarm, asymmetric tolerance band
- Three different power failure scenarios
- Logbook: records of every door opening, events, alarms and operating hours
- Language options EN, DE, FR, ES
- Wide range of program functions: Start after time, start after temperature, standstill segment
- Programs and temperature settings transferable to other HettCubes without additional software
- Interior made of high-quality stainless steel Special accessories:
- Shelfs and racks for storage of samples
- shelf for Petri Dishes
- tightly welded HTS drawers (various heights)
HettCube 400 incubators | HettCube 400 R cooled incubators | |
Temperature range | 1 K above ambient temperature up to +65 °C | 0 °C up to +65 °C |
Exterior dimensions (without bushing and door handle) W x D x H in mm | 710 x 825 x 1425 | 710 x 825 x 1425 |
Interior dimensions W x D x H in mm | 535 x 690 x 850 | 535 x 690 x 850 |
Internal volume in liters | 310 | 310 |
Validated usable volume in liters | 199 | 199 |
Percentage validated usable volume / internal volume | 64 % | 64 % |
Footprint in m² | 0,6 | 0,6 |
Weight in kg | 117 | 128 |
Number of trays provided as standard | 3 (2 standard + 1 HTS) | 3 (2 standard + 1 HTS) |
Temperature fluctuation at +37 °C | ± 0.1 K | ± 0.1 K |
Temperature uniformity at +37 °C | ± 0.2 K | ± 0.2 K |
Temperature uniformity at +25 °C | ± 0.1 K | ± 0.1 K |
Recovery time after door has been opened for 30 s at +37 °C | ≤ 4,5 min | ≤ 4,5 min |
Energy consumption at +37 °C | 0.046 kWh | 0.046 kWh |
Noise level | ≤ 41 dB (A) | ≤ 44 dB (A) |
Power supply | 220 – 240 V 1 ~ / 50 – 60 Hz | 220 – 240 V 1 ~ / 50 – 60 Hz |
Cat. No. | 64000 | 64005 |
Non-IvD version | 64001 | 64006 |
120 V 1 ~ / 50 – 60 Hz | 64000-01 | 64005-01 |
Weight in kg* | 122 | 133 |
Energy consumption at +37 °C* | 0.05 kWh | 0.05 kWh |
Noise level* | ≤ 42 dB (A) | ≤ 44 dB (A) |
For other products from Hettich Lab Technology, click here.